The fastest, most accurate, reliable, easy-to-use handheld water quality analysis tool for any pool or spa. POOLPRO™ does the job of up to nine tests in one easy to use instrument:
POOLPRO is a comprehensive high performance tool designed to simplify pool and spa water quality control for the pool professional. Both POOLPRO models feature innovative user friendly features and functions that make it easy to manage parameters critical to disinfection, water balance, system maintenance and compliance.
The Myron L® Company FCE Function reports FAC quickly and accurately by measuring ORP, the chemical characteristic of chlorine that directly reflects its effectivity, cross referenced with pH. Both DPD kits and colorimeters may tell the user the FAC value of the sample in the test tube, but since the chemistry of that sample is quite different from the source water being analyzed, the results are imprecisely related to actual disinfection power. The Myron L® Company FCE function measures the real, unaltered chemistry of source water, including moment-to-moment changes in that chemistry. FCE can be used for other types of oxidizing germicides and will track the effect of additives, such as cyanuric acid, that degrade chlorine effectivity without changing the actual concentration of free available chlorine present.
The PS9TKB adds the ability to perform in-cell conductometric titrations that provides a convenient way to determine alkalinity, hardness and LSI in the field. This eliminates the need to collect and transport samples to another location for analysis. User intuitive display prompts guide you through titration procedures from start to finish. All required reagents and equipment are included in the PS9 titration kit.
The POOLPRO PS9TKB features an LSI Calculator that allows you to calculate the LSI based on saved measurement values. This allows you to accurately calculate a saturation index value of a specific solution to determine whether the solution is balanced, scaling or corrosive. The LSI Calculator also allows you to adjust the measurement values to perform what-if scenarios to predict how changes in solution parameters would affect the water balance of a system.
The Hardness titration function allows you to set the hardness unit preference to either grains of hardness or ppm CaCO . The selected unit is also used in the LSI Calculator function.
Myron L® Company’s POOLPRO provides a fast, precise, easy-to-use method of obtaining Oxidation Reduction Potential (ORP or REDOX) mV readings to check the true level of effectiveness of ALL sanitizers in any pool or spa. ORP objectively and precisely measures sanitizer ability to burn up, or oxidize, organic matter in the water. ORP can only be determined by an electronic instrument.
POOLPRO ORP mV readings serve as a necessary check to ensure automatic ORP control systems are working properly. PoolPro also provides independent readings for recalibration and to detect system failure.
PoolPro provides a convenient one-touch test for Mineral/Salt concentration. This is ideal for saltwater systems where manual testing with separate instrumentation is necessary to ensure the proper amount of sodium chloride is present for chlorine generation in quantities specified for microbial disinfection. PoolPro can also be used to recalibrate equipment as part of regular maintenance.
The optional bluDock™ accessory package is an integrated data solution for your record keeping requirements, eliminating the need for additional hardware, wires and hassle. Because the user never touches the data, there is little opportunity for data tampering and human error. bluDock software, Guardian2 , has an easy to use interface with user intuitive functions for storing, sorting and exporting data.
includes PS9; cell extender; titration plunger; 100µL pipette; 12 tips; 3 buffers (pH 4, 7, and 10); pH/ORP Sensor Storage Solution; ORP Conditioner Solution; titration reagents H2SO4-1, HBUFF, EDTA-B-LC, EDTA-B-HC; Alkalinity Standard 100; Hardness Standard 200; standard solutions KCI-7000, 442-3000, and NaCl-7500.
All bottles are 2 oz/59 mL*. model: PS9TKB
PS9TKB AHL Titration Kit can be ordered with the bluDock Accessory Package, as well.
Display: 4 Digit Liquid Crystal Display |
Dimensions: LxWxH 196 x 68 x 64 mm/ 7.7 x 2.7 x 2.5 in. |
Weight: 352 g/12.4 oz. |
Case/conductivity cell material: VALOX* |
Cell capacities (w/o cell extender): pH/ORP: 1,2 mL/0.04 oz. Cond/TDS/MIN/SALT: 5 mL/0.2 oz. |
Power: 9V alkaline battery |
Battery life: >100 hours (5000 readings) |
Operating/storage temperature: 0 – 55°C/32 – 132°F |
Protection ratings: IP67/NEMA 6 |
Waterproof to: 1 meter/3 feet |
Conductivity | Ranges: 0–9999 μS/cm; 10–200 mS/cm; in 5 autoranges |
Resolution: 0.01(<100 μS); 0.1(<1000 μS); 1(<10 mS); 0.01(<100 mS); 0.1(<200 mS) | |
Accuracy: ±1% of reading | |
Auto Temperature Compensation: 0–71°C / 32–160°F | |
Conductivity/TDS Ratios Preprogrammed: KCl, 442*, NaCl | |
TDS | Ranges: 0–9999 ppm; 10–200 ppt; in 5 autoranges |
Resolution: 0.01(<100 ppm); 0.1(<1000 ppm); 1(<10 ppt); 0.01(<100 ppt); 0.1(<200 ppt) | |
Accuracy: ±1% of reading | |
Auto Temperature Compensation: 0–71°C / 32–160°F | |
Conductivity/TDS Ratios Preprogrammed: KCI, 442*, NaCl | |
Mineral/Salt | Ranges: 0–200 ppt; in 5 autoranges |
Resolution: 0.01(<100 ppm); 0.1(<1000 ppm); 1(<10 ppt); 0.01(<100 ppt); 0.1(<200 ppt) | |
Accuracy: ±1% of reading | |
Auto Temperature Compensation: 0–71°C / 32–160°F | |
Conductivity/TDS Ratios Preprogrammed: KCI, 442*, NaCl | |
pH | Ranges: 0–14 pH |
Resolution: 0.01 pH | |
Accuracy: ±0.01 pH** | |
Auto Temperature Compensation: 0–71°C / 32–160°F | |
ORP | Ranges: ±999 mV |
Resolution: 1 mV | |
Accuracy: ±1 mV | |
Free Chlorine FCE™ | Ranges: 0.00–9.99 ppm when 350≤ORPmV<725 and 0.0≤pH<9.9 and when 725≤ORPmV<825 and 0.0≤pH<8.9 |
Resolution: 0.01 ppm | |
Accuracy: <1.00ppm ±0.3ppm** and ≥1.00ppm ±0.2ppm | |
Auto Temperature Compensation: 0–71°C / 32–160°F | |
Alkalinity Titration | Ranges: 10-800 ppm |
Resolution: 0.1 ppm | |
Hardness Titration | Ranges: 0-1710 ppm / 0-100 grains |
Resolution: 0.1(<1000 ppm); 1(<1710 ppm); 0.1(<100 grains) | |
LSI Calculator | Ranges: -10 to +10 |
Resolution: 0.01 | |
Temperature | Ranges: 0–71°C / 32–160°F |
Resolution: 0.1°C/F | |
Accuracy: ±0.1°C |